State Primary Standard of Absorbed Dose Units, Absorbed Dose Rate of X-rays and Gamma Radiation

Reproduction and maintenance of the units of absorbed dose, power of absorbed dose of X-ray gamma radiations. Dissemination of the unit of absorbed dose, power of absorbed dose of X-ray gamma radiations to the working standards and working instruments of the measuring technology by means of the method of direct measurements and comparisons with a help of comparator in accordance with DSTU 4087-2001 “Metrology. National verification scheme for measuring instruments of the unit of absorbed dose, power of absorbed dose of X-ray gamma radiations”.

Metrological characteristics

Range

from 1·10-4 Gy to 10,0 Gy

from 1·10-10 Gy·s-1 to 1·10-3 Gy·s-1

Uncertainty by type A (uA)

(0,1–1,0)·10-2

Uncertainty by type B (uB)

(0,21–0,62)·10-2

Total standard uncertainty (uC)

(0,23–1,2)·10-2

Extended uncertainty (U)

(0,5–2,4)·10-2