State Primary Standard of Units of Exposure Dose, Power of Exposure Dose of X-ray and Gamma Radiation
Reproduction and maintenance of the units of exposition dose, power of exposition dose of X-ray and gamma radiations.
Dissemination of the units of exposition dose, power of exposition dose of X-ray and gamma radiations to the secondary standards, working standards and working instruments of the measuring technology by means of the method of direct measurements and comparisons with a help of comparator in accordance with DSTU 4089-2001 “Metrology. National verification scheme for measuring instruments of exposition dose, power of exposition dose of X-ray and gamma radiations”.
Metrological characteristics | ||
Range | from 3·10-11 C·kg-1 до 1·10-4 C·kg-1 from 3·10-12 A·kg-1 до 3·10-5 A·kg-1 | |
Uncertainty by type A (uA) | (0,1–1,0)·10-2 | |
Uncertainty by type B (uB) | (0,21–0,62)·10-2 | |
Total standard uncertainty (uC) | (0,23–1,2)·10-2 | |
Extended uncertainty (U) | (0,5–2,4)·10-2 |