State Primary Standard of Units of Exposure Dose, Power of Exposure Dose of X-ray and Gamma Radiation

Reproduction and maintenance of the units of exposition dose, power of exposition dose of X-ray and gamma radiations.

Dissemination of the units of exposition dose, power of exposition dose of X-ray and gamma radiations to the secondary standards, working standards and working instruments of the measuring technology by means of the method of direct measurements and comparisons with a help of comparator in accordance with DSTU 4089-2001 “Metrology. National verification scheme for measuring instruments of exposition dose, power of exposition dose of X-ray and gamma radiations”.

Metrological characteristics

Range

from 3·10-11 C·kg-1 до 1·10-4 C·kg-1

from 3·10-12 A·kg-1 до 3·10-5 A·kg-1

Uncertainty by type A (uA)

(0,1–1,0)·10-2

Uncertainty by type B (uB)

(0,21–0,62)·10-2

Total standard uncertainty (uC)

(0,23–1,2)·10-2

Extended uncertainty (U)

(0,5–2,4)·10-2